Overview
HK-3536B Automatic Cleveland Open Cup Flash Point Tester conforms to
ASTM D92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester.HK-3536B can measure the flash point and fire point of petroleum products,it is applicable to all petroleum products with flash points above 79℃ (175℉)and below 400℃(752℉)except fuel oils.
Features
- Electric ignition plug
- Touch screen operation,glass PT100 sensor
- Independent heating module,heating rate meets the standard strictly,equipped with detecting interface
- The apparatus can ignition, sweep, detect flash point, alarm, calculate and print storage results, air cooling, and detect atmospheric pressure automatically
- CANNON standard oil calibration
Technical parameters
- Rated voltage:AC220V±10% 50Hz
- Measuring range:90~400℃
- Heating power:700W continuously adjustable
- Ignition method:Automatic ignition
- Ambient requirement:Temperature:5~40℃ Humidity≤85%